TM-XD20DV,24DV Table Type Steam Sterilizers With Pulse-Vacuum System Specification General The unit disinfects and sterilizes the objects with saturated steam, by the means of pulsely exhausting the cold air to clear the efforts from which to the temperature,and make the objects completely...
+

Minimum quantity for "TM-XD20DV,24DV Table Type Steam Sterilizers With Pulse-Vacuum System" is 1.

Port: Shanghai,China
Payment
Share to

TM-XD20DV,24DV Table Type Steam Sterilizers With Pulse-Vacuum System

Specification

General
The unit disinfects and sterilizes the objects with saturated steam, by the means of pulsely exhausting the cold air to clear the efforts from which to the temperature,and make the objects completely drying by the way of vacuum-exhausting and jacket baking.By using the physical characteristics of the pressured saturated steam will releasing plenty of hot while condensing,the unit will keep the sterilizing objects wet and at high temperature in a given time ,which spoils the virus and the spores together, and good sterilizing efforts will be received.
Characterisctics
1.The sterilizing procedure is automatically controlled, it is not necessary to be supervised, and used conveniently.
2.At lease 3 or more times pulse-vacuum exhausting, which is applied to wrapped or unwrapped objects,hollow equipment.
3.Drying function.
4.With inner printer, which record the for the full procedure of the sterilization.
5.With Bowie&Dick testing function, which is for the steam penetrating testing.
6.Self-testing malfunction, easy to maintain.
7.Safety functions        
♦Safety locking:
The steam can not enter the chamber and the unit will alarm when the door is not closed tightly,and the door can not the open when the inner pressure is exceed 0.027mpa.
♦Over pressure protection:
If the pressure exceed the working pressure, the safety valve will exhausting the air.
♦Low water protection:
The unit can not be started if the water don’t reach the high level,and the power will cut off directly.
♦Over current protection:
The broker will cut off the power by the over current happened.   

Chamber Volume 20 Litersφ250×350 mm 24 Litersφ250×430 mm
Max working temperature 134±1 134±1
Max working pressure 0.23mpa 0.22mpa
Adjusted sterilizing temperature range55~134 55~134
Adjusted  timer range 1~99mins1~99mins
Adjusted drying time range 1~99mins 1~99mins
Vacuum limit -0.08mpa-0.08mpa
Hot Average ≤2 ≤2
Power supply AC220V/50HZAC220V/50HZ
Dimension 67*50*45cm 67*50*45cm
Transport dimension70*58*60cm 80*58*55cm
Gross/Net Weight 54kg/47kg60kg/50kg