TM-XD50DV Table Type Steam Sterilizers With Pulse-Vacuum System

Write a review
TM-XD50DV
In stock
Price: click login to view
Get a quotation upon your custom quantity
TM-XD50DV Table Type Steam Sterilizers With Pulse-Vacuum System General The unit disinfects and sterilizes the objects with saturated steam, by the means of pulsely exhausting the cold air to clear the efforts from which to the temperature,and make the objects completely drying by the way of...
+

Minimum quantity for "TM-XD50DV Table Type Steam Sterilizers With Pulse-Vacuum System" is 1.

Port: Shanghai,China
Payment
Share to

TM-XD50DV Table Type Steam Sterilizers With Pulse-Vacuum System

General

The unit disinfects and sterilizes the objects with saturated steam, by the means of pulsely exhausting the cold air to clear the efforts from which to the temperature,and make the objects completely drying by the way of vacuum-exhausting and jacket baking.By using the physical characteristics of the pressured saturated steam will releasing plenty of hot while condensing,the unit will keep the sterilizing objects wet and at high temperature in a given time ,which spoils the virus and the spores together, and good sterilizing efforts will be received.

Characterisctics

1.The sterilizing procedure is automatically controlled, it is not necessary to be supervised, and used conveniently.
2.At lease 3 or more times pulse-vacuum exhausting, which is applied to wrapped or unwrapped objects,hollow equipment.
3.Drying function.
4.With inner printer, which record the for the full procedure of the sterilization.
5.With Bowie&Dick testing function, which is for the steam penetrating testing.
6.Self-testing malfunction, easy to maintain.

7.Safety functions.

♦Safety locking:

 The steam can not enter the chamber and the unit will alarm when the door is not closed tightly,and the door can not be open when the inner pressure is exceed 0.027mpa.

♦Over pressure protection:

 If the pressure exceed the working pressure, the safety valve will exhausting the air.

♦Low water protection:

 The unit can not be started if the water don’t reach the high level,and the power will cut off directly.

♦Over current protection:

The broker will cut off the power by the over current happened.   

Parameter

Chamber Volume50L
Max working temperature134 ℃ ±1 ℃
Max working pressure0.22mpa
Adjusted sterilizing temperature range55 ℃〜134℃
Adjusted  timer range1〜99mins
Adjusted drying time range1〜99mins
Vacuum limit-0.08mpa
Hot Average≤2 ℃
Power supplyAC220V / 50HZ
Dimension770×600 ×540mm
Transport dimension880×730 ×700mm
G.W/N.W130 / 106 kg