ICU Ventilator AG-HXJ01

Write a review
AG-HXJ01
In stock
Price: click login to view
Get a quotation upon your custom quantity
ICU Ventilator AG-HXJ01 Introduction 10.4” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform.  Recycle breathing pipe, ensure easy operating and keep tidy. Multiple working modes such as volume control and pressure limit, adapt to wide range patient....
+

Minimum quantity for "ICU Ventilator AG-HXJ01" is 1.

Port: Shanghai,China
Payment
Share to

ICU Ventilator AG-HXJ01

Introduction

  • 10.4” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform. 
  • Recycle breathing pipe, ensure easy operating and keep tidy.
  • Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
  • Vaporizer with temperature, flow compensation and self-lock function, keep safety anytime.
  • Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects;
  • Real time pressure-time, flow-time loop graphics and high precision O2 concentration detection function included.
  • Venturi effect technology ensure the machine working without air gas source.

 Safety

  • Three level alarming system, visual and sound alarm information.
  • With lots of alarming, reminding and protection functions.
  • Advanced power management control technology.
  • With built-in backup power source, when outside power source goes off, back-up power source starts to work.
  • Self-check before running, eliminate system mistake.
  • Separate design of electric and gas, keep safety running of ventilator.

 Specifications

           

Ventilation modes:

IPPV; A/C; SIMV; SPONT/CPAP; MANU

Ventilator parameter ranges

 

Tidal volume(Vt)

0, 20 mL ~ 1500 mL

Frequence(Freq)

1bpm ~ 100 bpm

I/E

4:1~1:8

PEEP

0cmH2O ~ 20 cmH2O

Pressure triggering sensitivity(PTr)

-20 cmH2O ~ 0 cmH2O(Based on PEEP)

SIGH

0(off)  1/100 ~ 5/100

Apnea Ventilation

OFF,5 s ~ 60 s

Monitored Parameters

 

Frequence(Freq)

0 /min ~ 100 /min

Tidal volume(Vt)

0 mL ~ 2000 mL

MV

Airway pressure

0 L/min ~ 99 L/min

0cmH2O  100cmH2O

Oxygen concentration

15 % ~ 100 >#/b###

 

Graphical display: 

 

P-T(pressure – time)

 

F-T(flow – time)

 

 

Size

1.Wooden case packing size(main engine):L 560*W 560*H 615mm , GW:35KG ; NW:17KG

2.Wooden case packing size(trolley):L650*W690*H 520 mm , GW:38KG NW:18G

 

Alarm and protection  

The AC power failure alarm

Power failure or no connection

Internal battery backup low voltage alarm

<11.3±0.3V

No tidal volume

No tidal volume within 6s

High oxygen concentration alarm

Low oxygen concentration alarm

19%-100>#/b###

18%-99>#/b###

High Airway  pressure alarm

Low Airway  pressure alarm

High Minute Volume alarm

Low Minute Volume alarm

Continuous Pressure alarm

20cmH2O-100cmH2O

0cmH2O-20cmH2O

1L/min-99L/min

0-20L/min

(PEEP+1.5kPa) over 16s

Suffocation warning

5s-60s no spontaneous ventilation

 

Working conditions

 

Gas source

O2,

Pressure

280kPa-600kPa

Voltage

-220V±22V

Power frequency

50Hz±1Hz

Input power

40VA

No reviews found