ICU Ventilator AG-HXJ01
ICU Ventilator AG-HXJ01
Introduction
- 10.4” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform.
- Recycle breathing pipe, ensure easy operating and keep tidy.
- Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
- Vaporizer with temperature, flow compensation and self-lock function, keep safety anytime.
- Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects;
- Real time pressure-time, flow-time loop graphics and high precision O2 concentration detection function included.
- Venturi effect technology ensure the machine working without air gas source.
Safety
- Three level alarming system, visual and sound alarm information.
- With lots of alarming, reminding and protection functions.
- Advanced power management control technology.
- With built-in backup power source, when outside power source goes off, back-up power source starts to work.
- Self-check before running, eliminate system mistake.
- Separate design of electric and gas, keep safety running of ventilator.
Specifications
Ventilation modes: | IPPV; A/C; SIMV; SPONT/CPAP; MANU |
Ventilator parameter ranges |
|
Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
Frequence(Freq) | 1bpm ~ 100 bpm |
I/E | 4:1~1:8 |
PEEP | 0cmH2O ~ 20 cmH2O |
Pressure triggering sensitivity(PTr) | -20 cmH2O ~ 0 cmH2O(Based on PEEP) |
SIGH | 0(off) 1/100 ~ 5/100 |
Apnea Ventilation | OFF,5 s ~ 60 s |
Monitored Parameters |
|
Frequence(Freq) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2000 mL |
MV Airway pressure | 0 L/min ~ 99 L/min 0cmH2O ~ 100cmH2O |
Oxygen concentration | 15 % ~ 100 >#/b###
|
Graphical display: |
|
P-T(pressure – time) |
|
F-T(flow – time) |
|
Size 1.Wooden case packing size(main engine):L 560*W 560*H 615mm , GW:35KG ; NW:17KG 2.Wooden case packing size(trolley):L650*W690*H 520 mm , GW:38KG NW:18G
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <11.3±0.3V |
No tidal volume | No tidal volume within 6s |
High oxygen concentration alarm Low oxygen concentration alarm | 19%-100>#/b### 18%-99>#/b### |
High Airway pressure alarm Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm | 20cmH2O-100cmH2O 0cmH2O-20cmH2O 1L/min-99L/min 0-20L/min (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
Working conditions |
|
Gas source | O2, |
Pressure | 280kPa-600kPa |
Voltage | -220V±22V |
Power frequency | 50Hz±1Hz |
Input power | 40VA |