AG-AM003 Anesthesia Machine
AG-AM003 Anesthesia Machine
Specification
Application
The Anesthesia Machine makes a good performance in intensive Care Units(ICU) , operation room, anesthesiology department, and other departments. From high to low acuity, simple to complex cases, pediatric to geriatric patients, We offer you wide choices that you need in ventilation, monitoring and technique.
Good performance based on high security, high accuracy, high stability and accurate monitoring.
Features
1. Simplicity: four static casters with self-locking function.
2. Precision in an anesthesia ventilator with multiple ventilation modes: IPPV ,A/C,SIMV, and MANU
3. 10.4"TFT LCD screen displays the Ventilation parameters, alarm information and Oscillographs
4. A vaporizer with temperature, flow compensation and self-lock function, keep safety anytime
5. Pressure time low-time and high precision ETCO2, O2 concentration show in real-time
6. ETCO2 and Anesthesia Gas Scavenging System(AGSS)are optional
7. Built-in backup battery provides the emergency power supply to the unit
8. Low O2 pressure alarm and N2O, cut-off protection
9. Over 5.000 units are installed in more than 200 countries
Ventilation modes: | IPPV; A/C; SIMV; SIGH; |
Ventilator parameter ranges | / |
Flow meter:1 | O2(0.1-10L/min) N2O(0.1-10L/min) AIR(0.1-10L/min) |
Rapid oxygen supply | 25L/min-75L/min |
Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
Frequence(Freq) | 1bpm ~ 100 bpm |
I/E | 2:1-1:6 |
PEEP | 0cmH2O ~ 30 cmH2O |
Pressure triggering sensitivity(PTr) | -20 cmH2O ~ 0 cmH2O(Based on PEEP) |
Flow trigger sensitivity (FTr) | 0.5 L/min ~ 30 L/min |
SIGH | 0(off) 1/100 ~ 5/100 |
Apnea Ventilation | OFF,5 s ~ 60 s |
Pressure Limit | 20 cmH2O ~ 100 cmH2O |
Monitored Parameters | / |
Frequence(Freq) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2000 mL |
MV | 0 L/min ~ 100 L/min |
Oxygen concentration | 15 % ~ 100% |
Oscillographs display: | P-T(pressure-time) F-T(flow-time) |
Alarm and protection | / |
The AC power failure alarm | Power failure or no connection |
Low voltage alarm for backup battery | <11.3 ±0.3V |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm | 19%-100% |
Low oxygen concentration alarm | 18%-99% |
High Airway pressure alarm | 20cmH2O-100cmH2O |
Low Airway pressure alarm | 0cmH2O-20cmH2O |
High Minute Volume alarm | Adult(5L/min-20L/min) Paed(1L/min-15L/min |
Low Minute Volume alarm | 0-10L/min |
Continuous Pressure alarm | (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error | Show on screen |
Oxygen deficit | Show on screen |
Working conditions | / |
Gas source | O2, N2O,AIR |
Pressure | 280kPa-600kPa |
Voltage | 100-240V |
Power frequency | 50/60Hz |
Input power | 80VA |