Auto Hematology Analyzer BK-3100
Auto Hematology Analyzer BK-3100
Features:
1. 9.7-inch color touch LCD screen, easy to operate, display all parameters and histograms on the same screen
2. Fewer samples are required for testing: whole blood≤10ul, pre-diluted≤20ul
3. 60 samples/hour.
4. 21 parameters + 3 histograms
5. It can be turned on continuously for 24 hours
6. Automatically clean the sampler, counting pool and pipeline after starting up
7. Built-in thermal printer, the external printer can be connected.
8. Electrical impedance method, cyanide-free detection of HGB
9. Fully intelligent blockage removal and alarm function
10. Both automatic calibration and manual calibration are available
11. Counting Mode: Venous, capilary, prediluted
12. Support the LIS system
Specification
Model | BK-3100 |
Throughput | 60 samples/hour |
Principle | Impedance for WBC differentiation and WBC/RBC/PLT count; Colorimetric method for HGB |
Channels | 2 |
Parameters | 21 parameters(including WBC, Mid#, Lym#, Gran#, Mid%, Lym%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV,PDW,PCT, P-LCR, P-LCC) |
Histograms | 3 histograms (including WBC, RBC, PLT histograms) |
Sample Volume | Whole blood: 10μl, Prediluted: 20μl |
Calibration | Manual, auto calibration |
Counting Modes | Venous, capilary, prediluted |
Printing Model | Auto print, manual print |
LIS | Support Bi-directional LIS |
Language | Chinese, English, Spanish, French, Russian, Ukrainian |
Reagent | Diluent 20L, Hemolytic agent 500ml, Probe cleanser: 50ml |
Data Input | 9.7-inch touch screen, mouse, keyboard (optional) |
Output | Internal printer, support external printer |
Printer paper | 57*30mm |
Interface | 4 USB ports, network port,RS232 serial port. |
Storage | 100,000 results with histogram |
Unclog | High voltage burn, reverse high-pressure flush |
Work Temperature | 15-30°C |
Power Supply | 100~240 AC, 50/60Hz, 150VA |
Relative Humidity | 30﹪~85﹪ |
Atmospheric Pressure | 70.0kPa-110.0kPa |
External Size(W*D*H) | 295*467*418mm |
Packed Size(W*D*H) | Instrument: 404*564*616mm; Reagent: 480*480*480mm |
Net weight | 19 KG |
Gross weight | Instrument: 31KG; Reagent::25Kg |
Carry-over rate
Parameter | CV |
WBC | ≤ 1.5% |
RBC | ≤ 1% |
HGB | ≤ 1% |
PLT | ≤ 3% |
Background
Parameter | Background |
WBC | ≤ 0.3×109/L |
RBC | ≤ 0.03×1012/L |
HGB | ≤ 1 g/L |
HCT | ≤ 0.5% |
PLT | ≤ 10×109/L |
Linearity
Parameter | Measurement range | CV |
WBC | (0.5~10.00×109)/L (10.10~99.9×109)/L | (±0.5×109)/L ±5% |
RBC | (0.05~1.00×1012)/L (1.01~9.99×1012)/L | (±0.05×1012)/L ±5% |
HGB | (2~70g)/L (71~300.0g)/L | (±2.0g)/L ±2% |
PLT | (10~100×109)/L (101~999×109)/L | (±10.0×109)/L ±8% |
Precision
Parameter | Whole blood (CV) | Pre-diluted (CV) | Measurement range |
WBC | ≤ 2.0% | ≤ 4.0% | (4.0-15.0×109)/L |
RBC | ≤ 1.6% | ≤ 3.0% | (3.5-6.0×1012)/L |
HGB | ≤ 1.6% | ≤ 3.0% | (110.0-180.0g)/L |
MCV | ≤1.0% | ≤2.0% | (70.0-120.0)fL |
PLT | ≤ 4.0% | ≤8.0% | (150.0-500.0×109)/L |