Hospital Double-Channel Infusion Pump AG-CX-XB-Y1200
Hospital Double-Channel Infusion Pump AG-CX-XB-Y1200
AG-XB-Y1200 Double-Channel Infusion Pump comes with an LCD display and microcomputer controller. The peristaltic pump is a power source with multiple sensors to monitor the infusion pump and has a variety of alarm functions. It's easy to operate and meets all the demands of various cases of infusion. There are two infusion channels, this infusion pump can work the two channels simultaneously, and also can work with one of the two.
Functions
Automatic infusion: Set the infusion volume and speed. The peristaltic pump which is controlled by a microcomputer will automatically infusion according to the setting infusion speed.
KVO status: After finishing the total volume of infusion, the pump will switch to KVO status automatically. (keep vein open status).
Audible and visual alarm: Five kinds of alarms including Occlusion alarm, Bubble alarm, Door Open Alarm, Infusion Finished alarm, and Under-voltage alarm. When infusion failed, audible and visual alarms will occur and remind the operators to dispose of in time.
Apply to a variety of liquids: It can be used to infuse colorless transparent liquid and high nutrient solutions and color opaque liquid. Such as blood plasma etc, which is not recommended for infusing.
Suitable IV set: Ordinary IV set: Normal PVC IV set (pipe diameter is about 3.5mm); Make sure the pipe wall of the IV set must have an elasticity coefficient. Before using the new brand IV set should be calibrated.
Note: Other specifications of the IV set may not be able to maintain accuracy (even if the precision was calibrated) Special IV set: special IV set using high elastic silicone tube, please contact our company for purchasing.
AC and DC power supply: Built-in Li-ion rechargeable battery pack, when AC power is off, the system will switch to battery power supply immediately. when battery volume is insufficient the charging will automatically start, when battery volume is full the charging will automatically stop (Not less than 4 hours ).
Bolus mode: Press the bolus key twice quickly and enter into bolus mode, when the infusion pump is not running, the bolus liquid will not be counted into infused volume, when the infusion pump is running, the bolus liquid will be counted into infused volume. Until the release of the bolus key, the bolus mode will stop.
Infusion rate: It has two infusion rate modes: drop/min and ml/h.
Note: drop/min and ml/h are conversed by 20 drops/ml, which is different from the actual IV set.
Specification
Model | AG-XB-Y1200 Hospital Double-Channel Infusion Pump |
Flow Rate Range | 0.1-1200Ml/h |
Tube Type | Matched Tube Type Match all types(PVC), The special tube should contact supplier |
Adj. Infusion Range | 0-9999 |
Infusion Precision | ±3%(Adj.) |
KVO Speed | 1-3Ml/h |
Alarm | Beep and Led Alarms Power switch /occlusion /air in tube /door open /infusion finished /low battery /no operate /tube error installed / no medicine liquid /tube spoilage |
Drop Ctl Precision | No |
Drop Scan Sensor | No |
Sensor | Air Scan Sensor Ultrasonic |
Pressure Range | 30Kpa-160Kpa (High, Middle, Low Adj.) |
Beep Adj.Range | High, low |
Ac Power | AC100V-240V±10% 50Hz-60Hz |
Battery | Insider Battery lithium battery Infusion time≈4hours with battery |
Lithium cell(12V) | Time≈6-12hours with battery |
Power consumption | Less than 18VA |
Security Classification | Class I, BF type |
Working Condition | Temperature±5°C-+40°C Humidity 20%-90>#/td### |
Purge | Yes |
Ml/h | drop/min Auto Change |
Dimension | 188*115*196mm |
Weight (kg) | 3.8 |