Anaesthetic Workstation MDW-AKB4
Anaesthetic Workstation MDW-AKB4
Main body | High-strength engineering plastic rack, light, beautiful, and corrosion resistant |
Scope of application | adult, child |
Gas source | O2:0.28-0.6MPa N2O:0.28-0.6MPa |
Flow meter | O2:0.05-1.0L/min 1.1-10L/min N2O:0.05-1.0L/min 1.1-10L/min |
O2, N2O linkage, and N2O Stopper | When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure is <0.2Mpa, the flow of nitrous oxide would be cut off |
Flow mate of rapid oxygon supply | 25-75L/min |
Low oxygen pressure alarm | It will be alarmed when the oxygen pressure is <0.2MPa |
Vaporizer | It has the function of automatic compensation based on pressure-temperature and flow rate. The regulation range of evaporator concentration is 0-5vol%. Among Halothane, Enflurane, Isoflurane, and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
Respiratory circuit | Working mode:all-close,semi-close,semi-open APL:≤12.5kPa |
Respiratory Bellows | Bellows for adults, bellows for children Tidal volume range:0-1500ml |
Ventilator
Display mode | High-definition 5.7 LCD screen display |
Ventilation mode | IPPV, SIPPV, SIMV, MANUAL |
Ventilation function | PEEP, SIGH, IRV |
Ventilation parameters | Tidal volume: 50-1500ml Rate: 2-99bpm SIMV rate: 2-20bpm I:E: 2:1-1: 8 Inspiratory trigger pressure: -10-20cmH2O PEEP: 2-20cmH2O Pressure limitation: 10-60cmH2O SIGH: 1.5 times the inspiratory time/100bpm |
Parameters for ventilation monitoring | Tidal volume, Minute Ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure-time waveform, flow rate-time waveform, PEEP, inspiratory trigger pressure, average pressure, oxygen concentration, battery capacity |
Security Alert System | Oxygen Concentration alarm, Airway pressure alarm, Minute ventilation volume alarm, Tidal volume, Sustained high-pressure alarm, Suffocation alarm, Power alarm, Intubation off the alarm, battery voltage alarm |
Oxygen Concentration alarm | Upper limit setting range: 21%-99% Low limit setting range: 18%-80% |
Airway pressure alarm | Upper limit setting range: 5-60cmH2O Low limit setting range: 0-50cmH2O |
Minute ventilation volume alarm | Upper limit setting range: 1~20L/min Low limit setting range: 0~19L/min |
Tidal volume | Upper limit setting range: 100-1500ml Low limit setting range: 0-1400ml |
Sustained high-pressure alarm | It will give an alarm when stress has consistently been higher than 25cmH2O |
Power | AC 220V 50Hz (UPS, Storage battery) |
Function | Mechanical Arms |
Optional Items | End-tidal CO2 monitoring Specific Interface of waste anesthesia gas |