SZAS-IP-900 Infusion Pump Features: ●4.3" color LCD screen, backlight display ●Three working modes: Rate/Time/Volume mode ●Purge, KVO function ●Heating function ●Remote controller function ●Audible and Visible alarms ●Alarm: Infusion Complete, Empty, Faulty Signal, Misoperation, Occlusion,...
+

Minimum quantity for "SZAS-IP-900 Infusion Pump" is 10.

Port: Shanghai,China
Payment
Share to

SZAS-IP-900 Infusion Pump

Features:

●4.3" color LCD screen, backlight display
●Three working modes: Rate/Time/Volume mode
●Purge, KVO function
●Heating function
●Remote controller function
●Audible and Visible alarms
●Alarm: Infusion Complete, Empty, Faulty Signal, Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, ldle
●Calibration function, suitable for any brand of disposable IV set
Standard: Heating function, Drop sensor, Remote controller

Technical Specification

  • Pump Type : Drip rate type infusion pump
  • Mechanism : Peristaltic pump
  • Adjustable Drop Rate Range: 1- 400 drops/min (step: 1 drop/min)
  • Adjustable Volume Rate Range: 1-1200 mI/h (when 1-99.9 ml/h, step: 0.1 m/h; greater than 99.9 ml/h, step: 1 ml/h)
  • Flow Rate Accuracy: Within ±3% (using calibrated high quality IV set)
  • Mechanical Precision: within ±2%
  • Purge Rate: 800 ml/h
  • KVO Rate: 1 mlh (1- 300 mlh)
  •                 2 mlh (100~300mlh)
  •                 3 ml/h (greater than 300 ml/h)
  • Infusion Volume Range:1 ml-9999.9 ml (step: 0.1 mI)
  • Maximum Accumulated Volume: 9999.9 ml (step: 0.1 ml)
  • Time Range: 1 min-9999 min (step: 1 min)
  • Occlusion Alarm Threshold

High: 800 mmHg±200 mmHg (106.7kPa±26.7kPa)

Medium: 500 mmHg±100 mmHg (66.7kPa±13.3kPa)

Low: 300 mmHg+100 mmHg (40.7kPa±13.3kPa)

  • Alarm
Infusion Complete, Empty, Faully Signal, Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, Idle.
  • Air Bubble Detector
Method: ultrasonic wave, sensitivity: ≥225 μL
  • Fuse
F1AL/250V
2 pcs (installed inside pump).
  • Power Supply
AC 85~265V, 50/60Hz;
  • Internal Battery
11.1V rechargeable li-ion battery. Capacity:2000mAh; Over 4 hours' working when battery is charged for 8 hours, at flow rate of 25ml/h.
  • Power: ≤40VA
  • Size: 188 mm(L)*198 mm(W)*228 mm(H)
  • Weight: 2.2KG
  • Environment Requirements

Transpor

Ambient Temperature: -30°C~+55°C
Relative Humidity: 20%~95% (non-condensing)
Atmosphere Pressure: 70kPa~ 106kPa

Storage
Ambient Temperature: -30°C~ +55°C
Relative Humidity: 20%~95%
Atmosphere Pressure: 70~106 (kPa)

Operation
Ambient Temperature: +5~ +40°C
Relative Humidity: 20%~90%
Atmosphere Pressure: 86~106 (kPas)