SZAS-IP-900 Infusion Pump
SZAS-IP-900 Infusion Pump
Features:
●4.3" color LCD screen, backlight display
●Three working modes: Rate/Time/Volume mode
●Purge, KVO function
●Heating function
●Remote controller function
●Audible and Visible alarms
●Alarm: Infusion Complete, Empty, Faulty Signal, Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, ldle
●Calibration function, suitable for any brand of disposable IV set
Standard: Heating function, Drop sensor, Remote controller
Technical Specification
- Pump Type : Drip rate type infusion pump
- Mechanism : Peristaltic pump
- Adjustable Drop Rate Range: 1- 400 drops/min (step: 1 drop/min)
- Adjustable Volume Rate Range: 1-1200 mI/h (when 1-99.9 ml/h, step: 0.1 m/h; greater than 99.9 ml/h, step: 1 ml/h)
- Flow Rate Accuracy: Within ±3% (using calibrated high quality IV set)
- Mechanical Precision: within ±2%
- Purge Rate: 800 ml/h
- KVO Rate: 1 mlh (1- 300 mlh)
- 2 mlh (100~300mlh)
- 3 ml/h (greater than 300 ml/h)
- Infusion Volume Range:1 ml-9999.9 ml (step: 0.1 mI)
- Maximum Accumulated Volume: 9999.9 ml (step: 0.1 ml)
- Time Range: 1 min-9999 min (step: 1 min)
- Occlusion Alarm Threshold:
High: 800 mmHg±200 mmHg (106.7kPa±26.7kPa)
Medium: 500 mmHg±100 mmHg (66.7kPa±13.3kPa)
Low: 300 mmHg+100 mmHg (40.7kPa±13.3kPa)
- Alarm
- Air Bubble Detector
- Fuse
2 pcs (installed inside pump).
- Power Supply
- Internal Battery
- Power: ≤40VA
- Size: 188 mm(L)*198 mm(W)*228 mm(H)
- Weight: 2.2KG
- Environment Requirements
Transpor
Ambient Temperature: -30°C~+55°CRelative Humidity: 20%~95% (non-condensing)
Atmosphere Pressure: 70kPa~ 106kPa
Storage
Ambient Temperature: -30°C~ +55°C
Relative Humidity: 20%~95%
Atmosphere Pressure: 70~106 (kPa)
Operation
Ambient Temperature: +5~ +40°C
Relative Humidity: 20%~90%
Atmosphere Pressure: 86~106 (kPas)