Share to

SZAS-SP-500D Syringe Pump

Features

●4.3" color segment LCD screen, backlight display, can be used in various lighting conditions
●Simultaneous display: Time, Battery indication, Injection state, Mode, Speed, Injection volume and time, Syringe size, Alarm sound, Block, Accuracy, Body weight, Drug dose and liquid amount
●Speed, time, volume and drug amount can be adjusted via remote control, easier operation, save time of doctor and nurse
●Advanced technology, based on Linux system, more safe and steady
●Multi injection modes: Volume/Time/Body weight mode
●Visible and audible alarms cover all abnormal conditions

Technical Specification

Injection Rate Range

  • 50ml syringe : 0.1~999.9ml/h (step :0.1ml/h)    1000~1800ml/h (step : 1ml/h)
  • 30ml syringe :0.1~ 900ml/h (step :0.1ml/h)
  • 20ml syringe :0.1 ~ 600ml/h (step :0.1ml/h)
  • 10ml syringe :0.1~ 300ml/h (step :0.1ml/h)
  • 5 ml syringe :0.1~ 1 50ml/h (step :0.1ml/h)

Purge Rate

  • 50ml syringe: 1800ml/h (Bolus rate: 1200ml/h)
  • 30ml syringe: 900ml/h (Bolus rate: 600ml/h)
  • 20ml syringe: 600ml/h (Bolus rate: 400ml/h)
  • 10ml syringe: 300ml/h (Bolus rate: 200ml/h)
  • 5 ml syringe: 1 50ml/h (Bolus rate: 1 00ml/h)

Mechanical Precision
Within±2%

Preset Volume Range
0.1~1999.9ml
Occlusion Alarm Threshold
High: 800mmHg±200mmHg (106.7 kPa±26.7kPa)
Medium: 500mmHg±100mmHg (66.7 kPa±13.3kPa)
Lower: 300mmHg±100mmHg (40.7 kPa±13.3kPa)
KVO Rate
0.1~ 5ml/h (step: 0.1 ml/h)
Fuse
F1AL/250V
2 pcs (installed inside pump)

Power Supply
100~220V,50Hz, Built in Rechargeable Li-ion battery, 11.1V,Capacity≥2000mAh;After charging 10 hours, it
can work at 5ml/h (GB 9706.27-2005 medium rate) for 4 hours
Environment Requirement:
Operation:
a) Ambient temperature: +5°C~+40°C
b) Relative humidity: 20%~ 90%
c) Atmosphere Pressure: 86.0~ 106.0 (kPa)
Transportation & Storage:
a) Ambient temperature: -30°C~+55°C
b) Relative humidity: ≤95%
c) Atmosphere Pressure: 50.1~ 106.0 (kPa)

Alarm
Near, Injection finished, Preset dose injection finished, Occlusion, Syringe off, Syringe dislocated, Low battery, AC power off
Safe Class
Class I Type BF
Size: 282mm (L) x210mm (W) x 148mm (H)