AG-7300D Medical Diagnostic X-ray System,Digital Radiography Equipment
AG-7300D Medical Diagnostic X-ray System,Digital Radiography Equipment
Specification:
Power
Voltage | 380V±38V |
Frequency | 50Hz±1Hz |
Capacitance | ≥65kVA |
Internal resistance | ≤0.17Ω |
High Frequency High Voltage
X-Ray Generator
Power | 56KW |
Inverter frequency | 440 KHz |
Tube voltage | 40kV~ 150kV, continuous adjustment,1kv |
Tube current | 10mA~ 710mA, stepping adjustment |
Exposure time | 1.0ms~ 10000ms, stepping adjustment |
mAs range | 0.1 mAs~900 mAs |
X-ray source assembly (Toshiba E7252X)
Focal spot | 1.2mm /0.6mm |
Input power | 75kW/27kW |
Nominal tube voltage | 150KV |
Anode thermal capacity | 300 KHU |
Anode heat dissipation | 475W(667KHU/min) |
Tube thermal capacity | 1250KHU |
Rotary anode speed | 9700 rpm/min |
Collimator
Light | Halogen lamp,AC24V/100W |
Brightness | ≥100Lux |
Time | 30±2s |
Bucky stand
Up/down range | ≥1300 mm |
Grid | Density: 130 L/inch Ratio: 10:1 Focal distance: 180 cm Size: 15” x 18” |
Flat panel detector (Standard)
A / D transition | 14 bit |
Energy range | 40-150kVP |
Input power | DC24V |
Active area | 427(H)×427(V) |
Pixel matrix | 3072(H)×3072(V) |
Pixel pitch | 139 μm |
Limiting resolution | ≥3.1lp/mm |
Flat Penal Detector (Optional)
Toshiba FDX3543RP
Active area | 350(H)×430(V) |
Pixel matrix | 2448(H)×2984(V) |
Pixel pitch | 143 μm |
Limiting resolution | 3.7lp/mm |
A / D transition | 16 bit |
Energy range | 40 - 150 kVp |
Input power | DC 24V 2A |
Bed
Computer | Brand:DELL OptiPlex CPU: Intel I5 6500(3.20GHz) RAM: 4GB DDR3L Disk: 1T DVD drive: DVD-RW Net board: Broadcom NetXtreme 10/100/1000 PCIe Graphic Board: AMD Radeon R5 430,2G(DP,SL-DVI-I) |
Color LED display
Bed size | 2000*760mm |
Longitudinal movement | ≥900mm |
Transverse movement | ≥220mm |
Tube rotation | -90°~ +90° |
Longitudinal movement of the pillar | 1300mm |
Box loading specifications | 8”*10”~14”*17” |
bed SID |
Workstation soft
Basic operation: change the user password, edit user, acquire the image
Additional operation: add new checks, edit current checking information, add new positions, change image acquiring sequence, switch between multiple checking protocols, manual adjustment of exposure parameters, AEC, select focus, select patient’s body type, check tube heat content, select ESA curve, clip image, add remark (the remark will be sent to DICOM workstation), add position mark to image, rotate or flip image, view image in full size, check information of patient and exposure dose, accept or refuse image.
Image management: change list sequence order, edit patient’s basic information after exposure, check history image, resend history image, reprint history image, check additional image information, preview history image, manage refused image, space recycle
strategy, protect the image, manually delete the image
System management: edit user, change the user password, emergency registration set, check statistics information, detector calibration, equipment control, output queue management, image measurement.
No reviews found